- Volume Approx: 9 ft/ 256 L
- Interior Width: 25.2 in / 640 mm
- Interior Height: 31.4 in / 800 mm
- Interior Depth: 19.7 in / 500 mm
- Number of Stainless Steel Grids (Standard Equipment): 2
- Maximum Number of Grids: 9
- Maximum Loading per Grid: 44 lb / 20 kg
- Maximum Loading of Chamber: 661 lb / 300 kg
- Net weight approx.: 243 lbs.
- Gross weight carton approx.: 355 lbs
- 115 V (±10%), 50/60 Hz approx. 1800 W / 15.7 A
- 3 Year Warranty
Memmert UF260 Universal Forced Circulation Lab Oven, 115V | 9 Cu. ft.
Item#: UF260115VMemmert UF260 universal forced circulation laboratory oven is a 9 cu. ft. oven designed for precision heating, drying, conditioning, and material testing applications in research, pharmaceutical, and industrial laboratories. The unit provides temperature control from ambient +10°C to 300°C using a PID microprocessor controller and forced-air circulation technology to ensure excellent temperature uniformity throughout the chamber. A high-definition ControlCOCKPIT display supports advanced programming, calibration, and monitoring, while the stainless-steel interior offers durability and easy cleaning. Integrated overtemperature protection, auto-diagnostic capabilities, and a high-capacity chamber make the UF260 ideal for demanding thermal processing and large-volume laboratory workflows.
$7,101.00
REQUEST PROPOSAL- Temperature: 1 Pt100 sensor DIN class A in 4-wire-circuit
- Set temperature range: min.10°C above ambient up to +300°C
- Resolution of display for setpoint values 0.1°C up to 99.9°C, 0.5°C from 100°C and for actual values 0.1°C (LED)
- ControlCOCKPIT: adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-color display
- Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
- Adjustable Parameters: temperature (Celsius or Fahrenheit), air flap position, program time, time zones, daylight savings time
- Function SetpointWAIT: the process time does not start until the set temperature is reached
- Language Setting: German, English, Spanish, and French
- Calibration: three freely selectable temperature values
- Temperature control: adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
- Auto diagnostic system: for fault analysis