- Volume Approx: 26.4 ft / 749 L
- Interior Width: 40.9 in / 1040 mm
- Interior Height: 47.2 in / 1200 mm
- Interior Depth: 23.6 in / 600 mm
- Number of Stainless Steel Grids (Standard Equipment): 2
- Maximum Number of Grids: 14
- Maximum Loading per Grid: 66 lb / 30 kg
- Maximum Loading of Chamber: 661 lb / 300 kg
- Net weight approx.: 479 lbs.
- Gross weight carton approx.: 635 lbs
- 208 V, 3 phases w/o neutral, 50/60 Hz / 17.6 amps
- 3 Year Warranty
Memmert UF750 Universal Forced Circulation Lab Oven, 208V | 26.4 Cu. ft.
Item#: UF750Memmert UF750 universal forced circulation laboratory oven is a 26.4 cu. ft. oven designed for precision heating, drying, conditioning, and material testing applications in research, pharmaceutical, and industrial laboratories. The unit provides temperature control from ambient +10°C to 300°C using a PID microprocessor controller and forced-air circulation technology to ensure exceptional temperature uniformity throughout the chamber. A high-definition ControlCOCKPIT display supports advanced programming, calibration, and monitoring, while the stainless-steel interior offers outstanding durability, easy cleaning, and high sample capacity. Integrated overtemperature protection, auto-diagnostic capabilities, and a large-volume chamber make the UF750 an ideal solution for high-throughput thermal processing and large-scale laboratory workflows.
$14,496.00
REQUEST PROPOSAL- Temperature: 1 Pt100 sensor DIN class A in 4-wire-circuit
- Set temperature range: min.10°C above ambient up to +300°C
- Resolution of display for setpoint values 0.1°C up to 99.9°C, 0.5°C from 100°C and for actual values 0.1°C (LED)
- ControlCOCKPIT: adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-color display
- Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
- Adjustable Parameters: temperature (Celsius or Fahrenheit), air flap position, program time, time zones, daylight savings time
- Function SetpointWAIT: the process time does not start until the set temperature is reached
- Language Setting: German, English, Spanish, and French
- Calibration: three freely selectable temperature values
- Temperature control: adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
- Auto diagnostic system: for fault analysis