- Cu. Ft.: 16
- Door(s): One (1) swing glass door, self-closing, right hinged, not reversible, magnetic door gasket
- Exterior Dimensions: 25" W x 29" D x 79" H with casters
- Interior Dimensions: 20.375"W x 20.375"D x 48.75"H
- Shelves: Four (4) adjustable shelves (adjustable in 1/2" increments)
- Mounting and Installation: Swivel casters - locking front casters
- Interior lighting: LED interior lights are safety shielded and switch controlled
- Airflow Management: Non-applicable
- External probe access: 3/4" probe access port on rear wall
- Insulation: Cabinet is foamed-in-place with EPA compliant high-density urethane foam
- Exterior materials: White powder coated steel
- Access control: Keyed door lock; compatible with optional digital lock (Pyxis, Omnicell, and AcuDose RX)
- General warranty: Two (2) year parts and labor warranty
- Compressor warranty: Five (5) year compressor warranty
- Product Weight (lbs): 243
- Shipping Weight (lbs): 290
- Rated Amperage: 2 Amps
- Power Plug/Power Cord: NEMA 5-15 plug
- Facility Electrical Requirement: 110-120V AC, 15A breaker, NEMA 5-15 receptacle
- Agency Listing and Certification: ETL, C-ETL listed. Meets or exceeds ENERGY STAR v2.0 for Laboratory Refrigerators.
- Maximum temperature variation (Cabinet air): +/- 2.5° C
- Stability (Simulator ballast): +/- 0.5° C
- Adjustable Temperature Range: 36°F to 46°F (2°C to 8°C), Calibrated using a NIST traceable device, certificate included
- Stability (Simulator bag): Non-applicable
- Energy Consumption (KWh/day): <0.9 KWh/day *meets ENERGY STAR v2.0 for Laboratory Refrigerators
- Average Heat Rejection (BTU/hr): <250 BTU/h
- Noise Pressure Level (dBA): 40 dBA or less installed
- Pull down time to nominal operating temp: < 30 min (empty)
Performance Note: Max temperature variation based on an empty chamber at steady state using 15 bare (unweighted) thermocouple probes. Product loading, door openings, and ambient conditions will affect performance. Stability of simulator ballast based on bottle probe provided with the unit located center of empty chamber at steady state, no door openings